Determining the Envelope of the Blood Flow Velocity in Doppler Ultrasound

Authors

  • Jarosław Majewski DRAMIŃSKI S.A.
  • Krzysztof Opieliński Wrocław University of Science and Technology
  • Żogał Paweł CYBERMEDICS
  • Andrzej Wiktorowicz DRAMIŃSKI S.A.

Abstract

This paper describes and tests a method for automatically determining the envelope of the blood flow velocity curve in ultrasonic Doppler imaging, which was implemented in a prototype of the 128-channel mobile ultrasound B-mode scanner. On the basis of the determined envelopes, algorithms were also developed for the automatic determination of the most important characteristic points of the Doppler blood flow spectrum in pulse wave Doppler imaging mode and the most relevant blood flow parameters. Sufficiently good repeatability and precision were obtained with low computational complexity.